TIC20 Logtag

TIC20 Logtag 

Temperature Indicator with 20 day display

The LogTag® TIC20 is an irreversible electronic temperature 

indicator used to monitor temperature in pharmaceutical transport

applications of up to 20 days duration. 

The TIC20 offers an accurate and cost effective solution for 

monitoring the pharmaceutical cold chain in “last mile” distribution

 and other applications where full data logging capability is not 

required. Statistical data can be viewed directly on the device or 

downloaded using a dedicated LogTag® Interface Cradle and 

freely available LogTag® Analyzer software.

WHO PQS pre-qualified versions are also available.



คุณลักษณะ

Models TIC20
Operating Temperature Range -30°C ~ +60°C  (-22°F ~ +140°F)
Ambient humidity range during transport & use 0% to 95%RH
Resolution 0.1°C or better in range of -30°C to +60°C (0.2°F to -22°F, 1°F from 100°F to 140°F)
Accuracy Better than ±0.5°C for -10°C to +40°C (+14°F to +104°F)
Better than ±1.0°C for -30°C to -10°C (-22°F to +14°F)
Better than ±1.0°C for +40°C to +60°C (+104°F to +140°F)
Sensor Reaction Time T90 less than 7 minutes by method detailed in EN12830:199
Capacity Minimum & maximum temperature for each of the 20 days. 1st activation of each alarm for each day including trigger time and duration of each excursion. Please note the TIC20 does not store each individual temperature reading taken.
Sampling Frequency Factory set to 6 minutes
Memory type Non volatile
Recordings 20 day statistics, temperature/time excursions and alarm trigger time and duration for each day.
Start delay Factory set to 60 minutes
Sensor Precision electronic thermistor
Vibration Withstands vibration specification as detailed in EN12830:19
Shock Withstands shock specification as detailed in EN12830:1999. Withstand 5 drops 1m to smooth concrete floor without loss of function or calibration.
Environmental IEC 60529: IP64
Resistance to Electrical Storms Designed to resist the effects of intense electrical storms
Power source 3V Lithium-Manganese Dioxide, non-replaceable
Battery life Minimum storage life of 18 months before ‘start’. Monitoring period: 20 days. Minimum accessibility (display) period of 6 months after ‘stopped’
Size 73mm(W) x 34.0mm(H) x 8.6mm(T)
Weight 30g
Case Material Polycarbonate
Warranty One year replacement warranty against defects in materials and workmanship from the date of purchase. Excludes misuse or abuse of the product or as a result of unauthorized alteration or repairs
Calibration Certificate of Traceability and Calibration to ISO/IEC17025 available on request.
Alarm Events Single alarm event when recording temperatures lower than -0.5ºC for 60 minutes or longer.

มาตราฐานที่ได้รับ

EMC Compliance
EMC Compliance

EC EMC directives (EN 61000-6-1:2005 & EN 61000-6-3:2006) including electrostatic discharge as prescribed in EN 61000-4-2. Complies with FCC Part 15 Subparts A and B.

WHO Compliance
WHO/PQS Compliance

See TIC20 WHO PQS versions

RoHS Compliance
RoHS Directive

The RoHS Directive stands for "the restriction of the use of certain hazardous substances in electrical and electronic equipment". This Directive bans the placing on the EU market of new electrical and electronic equipment containing more than agreed levels of lead, cadmium, mercury, hexavalent chromium, polybrominated biphenyl (PBB) and polybrominated diphenyl ether (PBDE) flame retardants.